Factory Price Multi-Parameter ICU Portable Ambulance Hospital Monitor Machine

Factory Price Multi-Parameter ICU Portable Ambulance Hospital Monitor Machine

Categories: Patient monitor
Model NO.: MY-C003A

Functions:

Early warning of abnormal data
The hospital monitor machine can monitor and record ECG data for 24 hours, and automatically track and capture the patient's clinically valuable dynamic change data according to the patient's current ECG basic data and automatically store it.

Motion monitoring Multidimensional analysis
The hospital monitor machine realizes the monitoring of ECG signals in various human motion states, through the client software, remote data center analysis system and medical expert team for multi-level and multi-angle analysis and judgment, and gives users medical advice.

Touch screen operation is simple and convenient
The hospital monitor machine adopts a large-size touch screen design, which means that users can intuitively operate various functions through the screen, which is simple and convenient to use.

Shield the signal, accurate data
The hospital monitor machine can effectively shield the EMG signal and electromagnetic signal interference, ensure the accuracy of the ECG data and the validity of the analysis, and has clinical significance for the monitoring of abnormal heart conditions.

Quick details 

  • Place of Origin: CN
  • Instrument classification: Class II
  • Type:patient monitor
  • Brand Name:MAYA
  • Model Number:MY-C003A
  • Name:MY-C003A 7 inch patient monitor
  • parameters:ECG, HP, RESP, SPO2, NIBP
  • display:7 inch high resolution color TFT LCD display.
  • waveforms:Multi-lead ECG waveforms
  • battery:UP to 4 hours working capacity of built-in rechargeable battery.
  • usage:Suitable for adult,pediatrics and neonate patients.
  • function:Patient info input management function
  • capacity:Networking capacity.

Features :

5 Standard parameters:ECG, HP, RESP, SPO2, NIBP

Built-in lithium battery: 2000mAH

7 inch high resolution color TFT LCD display.

Real time S-T segment analysis, pace-maker detection.

Indication of Incorrect manipulation and malfunction analysis.

Patient info input management function Multi-lead ECG waveforms display in phase.

Large volume of tabular and graphic trends information storage and easy to recall.

Capture dynamic waveforms.

Efficient resistance to interference of defibrillator and HF knife.

UP to 4 hours working capacity of built-in rechargeable battery.

Unique accessory management cabinet. Networking capacity.

Rechargeable Lithium battery

Suitable for adult,pediatrics and neonate patients.

Standard Configuration: ECG, HP, RESP, SPO2, NIBP

Optional: ETCO2 Thermal recorder Wall mount Trolley

hospital monitor machinehospital monitor machine manufacturershospital monitor machinehospital monitor machine for sale

RESP

Measuring Method

RA-LL impedance

RESP Impedance Measuring Range 0.3~3Ω

Base Impedance Range 200~4000Ω

Bandwidth 0.1~2.5Hz

RESP Rate Range

Adult 0~120BrPM

Children and neonatal baby 0~150 BrPM

Resolution 1 BrPM

Precision 2 BrPM

Asphyxia Alarm 10~40 seconds

NIBP

Measuring Method Pulse wave oscillometry

Work Mode Manual/Automatic/STAT

Measuring Interval of Automatic Measuring Mode

1,2,3,4,5,10,15,30,60,90,120,180,240,480 minute(s)

Measuring Time of STAT Mode 5 minutes

PR range 40 – 240 bpm

Measuring Range and Precision

Range

AdultSystolic blood pressure40~270mmHg

Diastolic blood pressure10~215mmHg

Mean blood pressure 20~235mmHg

Children Systolic blood pressure40~200mmHg

Diastolic blood pressure10~150mmHg

Mean blood pressure 20~165mmHg

Neonatal baby Systolic blood pressure40~135mmHg

Diastolic blood pressure10~100mmHg

Mean blood pressure 20~110mmHg

Static pressure range 0~300mmHg

Static pressure precision3mmHg

Pressure precision:Max. average error: 5mmHg;Max.

standard deviation: 8mmHg

Overvoltage protection

Adult mode 300 mmHg

Children mode 240 mmHg

Neonatal baby mode150 mmHg

SpO2

Range 0~100%

Resolution1%

Precision 70~100%:2 DIGIT 0%~69%:no definition given

PR

Range 20~300bpm

Resolution1bpm

Precision 3bpm

ECG

Lead configuration

Standard 3-lead or 5-lead

3-leadRA,LA,LL,Lead method:I,II,III

5-leadRA,LA,LL,RL,V,Lead method:I,II,III,aVR,aVL,aVF,V

Increase 250,500,1000,2000

HR Range

Adult 15 ~ 300bpm (beat/minute)

Neonatal baby/children 15 ~ 350 bpm(beat/minute)

Precision±1% or ±1bpm,the larger prevails

Resolution 1 bpm(beat/minute)

Sensitivity > 200 uV (Peak-to-peak value)

Input Impedance > 5 (megohm)

Bandwidth

Diagnostic mode0.05~130Hz

Monitoring Mode 0.5~40Hz

Operation mode1~20Hz

Common Mode rejection Ratio

Diagnostic mode> 90 dB

Monitoring Mode > 100 dB

Operation mode> 100 dB

Pole Polarization Vvoltage Range 300Mv

Pacing Pulse Test

Test pacing pulse in accordance with the following conditions:

Amplitude:±2 mV ~ ±700mV

Width: 0.1ms ~ 2ms

Risetime: 10us~100μs

Pacing pulse inhibition

When pacing analysis switch is on, pacing pulse in accordance with the

following conditions are restrained, but affection against HR calculation.

Amplitude:±2 mV ~ ±700mV

Width: 0.1ms ~ 2ms

Risetime: 10us ~ 100μs

Baseline Recovering Time

After defibrillation< 3 seconds

Signal Range 8 mV (Peak-to-peak value)

Calibrating Signal

1mV(Peak-to-peak value),precision5%

ST Segment Measuring Volume

Measuring range: -2.0mV ~ +2.0mV

Measuring precision: Ranging -0.8 mV~+0.8mV, measuring error is

±0.02mV or ±10%,.

The larger prevails.

hospital monitor machineChina hospital monitor machinehospital monitor machine for sale

Certificate

.jpg




Send Message

*
*
* clear input
* clear input
* clear input
* clear input
customer-service Customer Service
whatsapp WhatsApp
wechat WeChat
message Message
go top Top
WeChat Close
qr code
18925044116
 

Send Message

*
*
* clear input
* clear input
* clear input
* clear input